Anesthesia Unit(A2200)
Model: A2200
Features:
1, CE and ISO certificated anesthesia machine
2, Two independent bellows separated for pediatric and adult, the working interface is independent.
3, Anesthesia ventilator with IPPV, SIMV, SIGH, MANUAL, Standby and Demo.
4, Compact CO2 absorber
5, Two high precious vaporizer, compatible with Penlon or Ohmeda vaporizer
6, Display: 10.4TFT color screen
Application: Adult, Pediatric
Driven mode: Pneumatically driven electronically controlled
Mode: Semi-open, Semi-closed or closed system
Anesthetic vaporizer: Two high precise vaporizers, Halothane, Enflurane or Isoflurane
Power supply: 220V± 10%, 50Hz± 1Hz
110V± 10%, 60Hz± 1Hz
Internal battery: DC12V( over 4 working hours)
Anesthesia Ventilator
Respiration mode:
IPPV, SIMV, SIGH, MANUAL, STANDBY, DEMO
Waveform: P-t, F-t
Tidal Volume: 50ml ~1600ml
Respiratory frequency: 1~65bpm
I: E ratio: 8: 1~1: 8
Inspiration pause: 0~55%Ti
Inspiration Pressure: 1~60CMH2O
Trigger Pressure: -20~20CMH2O
Monitoring parameters:
O2 concentration, MV, Respiratory frequency, Tidal volume, Ppeak, I: E ratio, Pmean, etc.
Pressure alarm setting:
High pressure alarm set 1~60CMH2O
Low pressure alarm set 0~20CMH2O
O2 concentration alarm setting:
High 21~99%
Low 21~50%
MV alarm setting:
MV high alarm set 10~99L/min
MV low alarm set 0~40L/min
Main Unit
Circle absorber: Compact circle absorber
Flow meter:
O2, N2O four-tube flow meter
O2 tube: 0L/min~1L/min~10L/min
N2O tube: 0L/min~1L/min~10L/min
O2 and N2O linkage device keep oxygen concentration over 25%
Oxygen flush: 25L/min~75L/min
Gas source pressure:
O2: 0.25MPa~0.65MPa
N2O: 0.25MPa~0.65Mpa
Gauge: 4 units, 2× N2O, 2× O2